Shaanxi Qindiyuan Environmental Protection Equipment Co., Ltd.
leonleesale@aliyun.com 86--15389206502
Products
Products
Home > Products > Industrial Sensors‌ > 2 Wire Programmable Transmitter PR 5334B High Accuracy With Galvanic Isolation Programmable Sensor Error Value

2 Wire Programmable Transmitter PR 5334B High Accuracy With Galvanic Isolation Programmable Sensor Error Value

Product Details

Place of Origin: Denmark

Brand Name: PR

Certification: CO

Model Number: PR 5334B

Document: 5334v111_uk.pdf

Payment & Shipping Terms

Minimum Order Quantity: 1

Price: 80usd

Delivery Time: 5-7 work days

Payment Terms: T/T

Supply Ability: 99pcs 4-6 weeks

Chat now
Highlight:

Programmable Transmitter High Accuracy

,

Galvanic Isolation Temperature Transmitter

Estimated Lead Time:
1-2 Weeks
Operating Voltage Range:
4.5 V DC ... 32 V DC
Compensation Rangeability:
Up To 400:1
Housing Material:
Stainless Steel
Using Range:
Temperature
Product Type:
Discrete Output Card
Aprroval:
None
Accuracies:
Up To 0.035% Standard & 0.025% Opt
Products Name:
Differential Pressure Transmitter
Temperature Compensation:
Automatic
Shipping Weight:
2kg
Input Protection:
Internally Fused, Non-replaceable
Country Of Origin:
United Kingdom And Thailand
Range:
-200℃-1200℃
Part Number:
1037778
Estimated Lead Time:
1-2 Weeks
Operating Voltage Range:
4.5 V DC ... 32 V DC
Compensation Rangeability:
Up To 400:1
Housing Material:
Stainless Steel
Using Range:
Temperature
Product Type:
Discrete Output Card
Aprroval:
None
Accuracies:
Up To 0.035% Standard & 0.025% Opt
Products Name:
Differential Pressure Transmitter
Temperature Compensation:
Automatic
Shipping Weight:
2kg
Input Protection:
Internally Fused, Non-replaceable
Country Of Origin:
United Kingdom And Thailand
Range:
-200℃-1200℃
Part Number:
1037778
2 Wire Programmable Transmitter PR 5334B High Accuracy With Galvanic Isolation Programmable Sensor Error Value
2-wire programmable transmitter
PR 5334B
Product Overview
  • TC input with high measurement accuracy
  • Galvanic isolation for enhanced safety
  • Programmable sensor error value
  • Designed for DIN form B sensor head mounting
Environmental Conditions
Operating temperature -40°C to +85°C
Calibration temperature 20...28°C
Relative humidity < 95% RH (non-condensing)
Protection degree (enclosure/terminal) IP68 / IP00
Mechanical Specifications
Dimensions Ø 44 x 20.2 mm
Weight approx. 50 g
Wire size 1 x 1.5 mm² stranded wire
Screw terminal torque 0.4 Nm
Vibration IEC 60068-2-6
2...25 Hz: ±1.6 mm
25...100 Hz: ±4 g
Electrical Specifications
Power Supply
Supply voltage 7.2...30 VDC
Internal power dissipation 25 mW…0.8 W
Isolation
Isolation voltage, test / working 1.5 kVAC / 50 VAC
Performance
Response time (programmable) 1...60 s
Voltage drop 7.2 VDC
Warm-up time 5 min.
Power on to stable output 4.5 s
Accuracy Better than 0.05% of selected range
Signal dynamics, input 18 bit
Signal dynamics, output 16 bit
Effect of supply voltage change < 0.005% of span / VDC
EMC immunity influence < ±0.5% of span
Extended EMC immunity: NAMUR NE21, A criterion, burst < ±1% of span
Input Specifications
Thermocouple Input
Thermocouple type B, E, J, K, L, N, R, S, T, U, W3, W5, LR
Cold junction compensation (CJC) < ±1.0°C
Sensor error detection Yes
Sensor error current: When detecting / else Nom. 33 μA / 0 μA
Voltage Input
Measurement range -12...150 mV
Min. measurement range (span) 5 mV
Input resistance 10 MΩ
Output Specifications
Current Output
Signal range 4…20 mA
Min. signal range 16 mA
Load (@ current output) ≤ (Vsupply - 7.2) / 0.023 [Ω]
Load stability ≤ 0.01% of span / 100 Ω
Sensor error indication Programmable 3.5…23 mA
NAMUR NE43 Upscale/Downscale 23 mA / 3.5 mA
Certifications & Approvals
Intrinsic Safety / Ex Marking
ATEX II 1 G Ex ia IIC T6...T4 Ga, II 2 D Ex ia IIIC Db, I M1 Ex ia I Ma
IECEx Ex ia IIC T6...T4 Ga, Ex ia IIIC Db, Ex ia I Ma
INMETRO Ex ia IIC T6...T4 Ga, Ex ia IIIC Db, Ex ia I Ma
Compliance Standards
EMC 2014/30/EU
ATEX 2014/34/EU
RoHS 2011/65/EU
EAC TR-CU 020/2011
EAC Ex TR-CU 012/2011
Approval Certificates
ATEX DEKRA 20ATEX0095X
IECEx DEK 20.0059X
INMETRO DEKRA 23.0009X
EAC Ex EAEU KZ 7500361.01.01.08756
DNV Marine TAA0000101
Programming requires PR 5909 Loop Link communications interface