Xi'an Yingzhuoke Industrial Equipment Co., Ltd.
leonleesale@aliyun.com 86--15389206502
Products
Products
Home > Products > Digital Multimeter DMM > Fluke 87V Industrial Multimeter

Fluke 87V Industrial Multimeter

Product Details

Place of Origin: USA

Brand Name: FLUKE

Model Number: Fluke 87V Industrial Multimeter

Document: bbe822ac-8c6a-4028-93b1-b10...le.pdf

Payment & Shipping Terms

Minimum Order Quantity: 1

Price: 600$-800$

Delivery Time: 1week

Payment Terms: T/T

Chat now
Highlight:

Fluke 87V digital multimeter

,

industrial multimeter with warranty

,

Fluke DMM for electrical testing

DC Voltage Range:
1000 V
AC Voltage Range:
1000 V
DC Current Range:
10 A
AC Current Range:
10 A
Resistance Range:
50 MΩ
Capacitance Range:
9,999 µF
DC Voltage Range:
1000 V
AC Voltage Range:
1000 V
DC Current Range:
10 A
AC Current Range:
10 A
Resistance Range:
50 MΩ
Capacitance Range:
9,999 µF
Fluke 87V Industrial Multimeter
Electrical Specifications
Measurement Parameter Specification
Voltage DC Maximum voltage 1000 V
Accuracy ±(0.05% + 1)
Maximum resolution 10 µV
Voltage AC Maximum voltage 1000 V
Accuracy ±(0.7% + 2) True RMS
AC bandwidth 20 kHz with low pass filter; 3 dB @ 1 kHz
Maximum resolution 0.1 mV
Current DC Maximum amps 10 A (20 A for 30 seconds maximum)
Amps accuracy ±(0.2% + 2)
Maximum resolution 0.01 µA
Current AC Maximum amps 10 A (20 A for 30 seconds maximum)
Amps accuracy ±(1.0% + 2) True RMS
Maximum resolution 0.1 µA
Resistance Maximum resistance 50 MΩ
Accuracy ±(0.2% + 1)
Maximum resolution 0.1 Ω
Capacitance Maximum capacitance 9,999 µF
Accuracy ±(1% + 2)
Maximum resolution 0.01 nF
Frequency Maximum frequency 200 kHz
Accuracy ±(0.005% + 1)
Maximum resolution 0.01 Hz
Duty Cycle Maximum duty cycle 99.9%
Accuracy ±(0.2% per kHz + 0.1%)
Maximum resolution 0.1%
Temperature Measurement Meter range (excluding probe) -200.0 °C to 1090 °C
-328.0 °F to 1994.0 °F
80 BK temperature probe -40.0 °C to 260 °C
-40.0 °F to 500 °F, 2.2 °C or 2% whichever is greater
Conductance Maximum conductance 60.00 nS
Accuracy ±(1.0% + 10)
Maximum resolution 0.01 nS
Diode Range 3 V
Resolution 1 mV
Accuracy ±(2% + 1)
Duty Cycle Range Accuracy Within ±(0.2% per kHz + 0.1%)
Environmental Specifications
Operating temperature –20 °C to +55 °C
Storage temperature –40 °C to +60 °C
Humidity (without condensation) 0% – 90% (0 °C – 35 °C)
0% – 70% (35 °C – 55 °C)
Operating altitude 2000 m
Safety Specifications
Overvoltage category EN 61010–1 to 1000 V CAT III, 600 V CAT IV
Agency approvals CE, CSA, RCM
Mechanical & General Specifications
Size 201 × 98 × 52 mm (with holster)
Weight 355 g
624 g – with holster
Display Digital: 6000 counts, updates 4/sec; 19,999 counts in high-resolution mode
Analog: 32 segments, updates 40/sec
Frequency: 19,999 counts, updates 3/sec at > 10 Hz
Warranty Lifetime
Battery life (alkaline) ~400 hours typical, without backlight
Shock 1 meter drop per IEC 61010–1:2001
Vibration Per MIL–PRF–28800 for a Class 2 instrument